Laboratory Medicine ›› 2018, Vol. 33 ›› Issue (6): 543-548.DOI: 10.3969/j.issn.1673-8640.2018.06.017

Previous Articles     Next Articles

Determination of reverse triiodothyronine by chemiluminescence assay

SONG Yunxiao1, WU Jianmin2, ZHANG Haichen1   

  1. 1. Department of Clinical Laboratory,Shanghai Xuhui Central Hospital,Shanghai 200031,China
    2. Department of Clinical Laboratory,Shanghai Yangpu District Mental Health Center,Shanghai 200093,China
  • Received:2018-03-10 Online:2018-06-30 Published:2018-07-06

Abstract:

Objective To evaluate the performance of reverse triiodothyronine(rT3) determination by chemiluminescence assay. Methods The precision,accuracy,analytical sensitivity,analytical specificity,linearity,reference range and methodological comparison of rT3 determination by chemiluminescence assay were evaluated according to EP documents of the Clinical and Laboratory Standards Institute(CLSI). The levels of 3,5,3'-triiodothyronine(T3),thyroxine(T4),thyroid-stimulating hormone(TSH) and rT3 were determined in 50 healthy subjects(healthy control group) and 131 patients,including 30 preliminarily diagnosed hyperthyroidism patients(hyperthyroidism group),28 hyperthyroidism patients with treatment for 1 month(hyperthyroidism treatment group),25 preliminarily diagnosed hypothyroidism patients(hypothyroidism group),23 hypothyroidism patients with treatment for 1 month(hypothyroidism treatment group) and 25 primary non-thyroid diseases [euthyroid sick syndrome (ESS)]. Results For high-level and low-level quality control materials of rT3,the within-run coefficients of variation(CV)were 4.82% and 4.95%,and the between-run CV were 7.51% and 5.90%,respectively. The biases were -1% and 1%. The linearity was 0.05-10.07 ng/mL,and the reference range was 0.20-0.95 ng/mL. There was a good correlation between chemiluminescence assay with radioimmunoassay(r=0.952). Compared with healthy control group,the levels of rT3,T3 and T4 in hyperthyroidism group were increased(P<0.05),and TSH was decreased(P<0.05). In hypothyroidism group,the levels of rT3 and T4 were decreased(P<0.05),and TSH was increased(P<0.05). Compared with hyperthyroidism group,hyperthyroidism treatment group had decreased rT3,T3 and T4 levels(P<0.05),and TSH was increased(P<0.05). Compared with hypothyroidism group,hypothyroidism treatment group had increased rT3 and T4 levels(P<0.05),and TSH decreased(P<0.05). The level of rT3 in ESS group was higher than those in hypothyroidism and healthy control groups(P<0.05),the levels of T3 and T4 were lower than those in healthy control group(P<0.05),there was no statistical significance compared with hypothyroidism group(P>0.05),TSH level was lower than that in hypothyroidism group(P<0.05),and there was no statistical significance compared with healthy control group(P>0.05). Conclusions The determination of rT3 by chemiluminescence assay has good precision,accuracy,specificity and linearity,which is suitable for large scale sample determination. It can be used for the differential diagnosis of hypothyroidism and ESS.

Key words: Reverse triiodothyronine, Chemiluminescence assay, Methodological evaluation

CLC Number: