Laboratory Medicine ›› 2018, Vol. 33 ›› Issue (1): 60-62.DOI: 10.3969/j.issn.1673-8640.2018.01.011
• Orginal Article • Previous Articles Next Articles
FENG Yuqing, HUANG Susu, LUO Jing, GU Guolong
Received:
Published:
Abstract:
Objective To study the consistency rate of UF-1000i automatic urine analyzer bacterial channel and mid-stream urine culturing,and to provide a reference for the early empirical use of antibiotics for urinary tract infection(UTI). Methods Urine specimens were determined for bacterial culturing and urinary sedimentation analysis. Bacterial culturing was performed by Vitek 2 Compact. Urinary sedimentation analysis was performed by UF-1000i becterial channel. Results Th. Results of mid-stream urine culturing were as gold standard. For UF-1000i bacterial channel,the consistency rate,sensitivity and specificity of Gram-positive coccus were 90.7%,60.8% and 93.5%,and those of Gram-negative bacillus were 93.3%,76.7% and 96.8% (P<0.05).Conclusions UF-1000i bacterial channel plays a role in the early clinical selection of antibiotics,especially in UTI caused by Gram-negative bacillus.
Key words: UF-1000i automatic urine analyzer bacterial channel, Mid-stream urine culturing, Consistency rate
CLC Number:
R446.65
FENG Yuqing, HUANG Susu, LUO Jing, GU Guolong. Consistency rate of UF-1000i automatic urine analyzer bacterial channel and mid-stream urine culturing[J]. Laboratory Medicine, 2018, 33(1): 60-62.