SYSMEX CS5100 血凝分析仪性能评估及凝血六项参考区间的调整 |
闫宏斌, 郑善銮, 郝晓柯 |
Performance evaluation and reference range adjustment for 6 coagulation indicators of SYSMEX CS5100 coagulation analyzer |
图2 CS5100与CA7000A2390的能力比对线性回归图 注:a图,PT, Y= 0 . 990 5 X+ 0 . 063 2, R 2 =0.997;b图,APTT, Y= 1 . 000 1 X- 1 . 416 9, R 2 =0.996 5;c图,FIB, Y= 0 . 966 9 X+ 9 . 110 3, R 2 =0.992 6;d图,TT, Y= 1 . 011 4 X- 1 . 539 2, R 2 =0.999 3;e图,DD, Y= 0 . 973 5 X- 0 . 176, R 2 =0.983 6;f图,FDP, Y= 0 . 953 8 X+ 0 . 905 5, R 2 =0.975 8 |
![]() |